Fluid Flow Rates: IV1000 STAT I.V. SET
Fluid flow rates (per minute and per hour) are provided for the IV1000 when suspended at various heights, with and without the addition of extension sets and various I.V. catheters
Test Conditions: The following fluid flow rates were determined with production run products. The IV-1000 sets tested were from Lot #01J8422; the TS-3002 from Lot #12H2965; the ET-3002 from Lot #06J1593; the ET-06L from Lot #11H2747; the Argyle, Medicut 12 ga. X 5 1/2″ IV catheter from Lot #434537; and, the Cook, Cook Critical Care, P. O. Box 489, Bloomington, IN 47402, 9.0 Fr. X 5 1/4″ IV catheter (Order No. CVSSN) from Lot #7375.37827. Tests were conducted with the IV apparatus attached to four (4), 5 liter Travenol collapsible fluid containers filled (and replenished as needed) with water. All heights were measured form the bottom of the lowest fluid container to the tip of the IV set, extension set or IV catheter. The fluid containers were suspended on an International WIN, Ltd. HG-7001 GYRO Fluid Hanger. When tests were conducted at 50 inches (4 feet, 2 inches) 2 fluid containers were on the low hooks and 2 on the higher hooks. Tests at 96 inches (8 feet) were conducted with all 4 fluid containers on the higher hooks. The fluid containers were interconnected with TS2001 Transfer Sets. All paired tests were conducted with a stop watch for 2 minutes. If the flow results for the system under test varied more than 5 ml they were repeated.
SYSTEM TESTED | ACTUAL FLOW (ML), 2 MIN |
FLOW (ML/MIN) |
FLOW (L/HR) |
IV-1000 STAT IV Set |
925 923 |
462 | 27.72 |
IV-1000 STAT IV Set ET-3002 Extension Set |
857 855 |
428 | 25.68 |
IV-1000 STAT IV Set ET-06L Extension Set |
686 687 |
343.25 | 20.56 |
IV-1000 STAT IV Set |
882 879 |
440.25 | 26.42 |
IV-1000 STAT IV Set |
817 816 |
408.25 | 24.50 |
SYSTEM TESTED | ACTUAL FLOW (ML), 2 MIN |
FLOW (ML/MIN) |
FLOW (L/HR) |
IV-1000 STAT IV Set |
826 829 |
413.75 | 24.83 |
IV-1000 STAT IV Set ET-3002 Extension Set Medicut 12 ga X 5 1/2″ |
786 786 |
393 | 23.58 |
IV-1000 STAT IV Set ET-3002 Extension Set Abbocath-T IV Catheter |
612 611 |
305.75 | 18.35 |
SYSTEM TESTED | ACTUAL FLOW (ML), 2 MIN |
FLOW (ML/MIN) |
FLOW (L/HR) |
IV-1000 STAT IV Set |
610 609 |
304.75 | 18.29 |
IV-1000 STAT IV Set ET-3002 Extension Set |
578 575 |
288.25 | 17.30 |
IV-1000 STAT IV Set ET-06L Extension Set |
477 480 |
239.25 | 14.36 |
IV-1000 STAT IV Set |
600 603 |
300.75 | 18.05 |
IV-1000 STAT IV Set |
559 560 |
279.75 | 16.79 |
IV-1000 STAT IV Set ET-3002 Extension Set Abbocath-T IV Catheter |
380 380 |
190.0 | 11.40 |
In addition, the number of drops per milliliter were determined in production IV-1000 STAT IV Sets, Lot #01J8422. Tests were conducted at 50 Inches (4 feet, 2 inches) and at 96 Inches (8 feet). Paired tests with 3 production sets were made. The number of drops were counted for 10 ml of fluid flow. Distances and test conditions were the same as for the foregoing tests.
SYSTEM TESTED | HEIGHT (INCHES) | DROPS/ML |
IV-1000 STAT IV Set | 50 | 12.8 |
IV-1000 STAT IV Set | 96 | 12.0 |
Tests Conducted by: William J. Donawick, DVM • July 5 and 26, 1988
Fluid Flow Rates: IV250 RAPID INFUSION SET
Fluid flow rates (per minute and per hour) are provided for the IV250 when suspended at two heights, with and without the addition of an I.V. catheter.
Test Conditions: The following fluid flow rates were determined with a production run IV250, Lot # 136047 and an Abbocath – T, 14G X 5 1/2” Teflon I.V. Catheter, Order No. 4535-84. Tests were conducted with the I.V250 attached to a single 5 liter flexible container of Abbott Normosol – R fluids. The quantity of fluids in the suspended container was depleted as the tests were conducted. Tests with the fluids suspended at 50” were done first. All tests were conducted with a stop watch for 2 minutes. Date: 4/12/2005.
SYSTEM TESTED | ACTUAL FLOW (ML), 2 MIN |
FLOW (ML/MIN) |
FLOW (L/HR) |
IV250 I.V. Set |
686 | 343 | 20.58 |
SYSTEM TESTED | ACTUAL FLOW (ML), 2 MIN |
FLOW (ML/MIN) |
FLOW (L/HR) |
IV250 I.V. Set |
1053 | 526.5 | 31.59 |
IV250 I.V. Set Abbocath-T I.V. Catheter |
480 | 240 | 14.4 |